Actual | YOCON automatic nucleic acid extraction workstation can effectively prevent cross-contamination!

Yocon Market

2021-10-20

Company Events


“Accurate, fast, convenient and secure”

This is a prioritization of our requirements for the workstation.

The workstation has to be put into operation.

One of the biggest concerns that customers often have about such product is that -

Whether the equipment will produce a false positive result (cross-contamination of different samples)?

To make customers more assured, we have tested the FE4800 automatic nucleic acid extraction workstation, using real virus samples (purchased from national institutions through regular channels, without infectivity), and have made the first success!


I. Background

What is the possible cause for cross-contamination of samples?

--Too short distance between sampling tubes and between different deep-well plates, and inevitable aerosol produced during liquid transfer by equipment. Cross-contamination of samples may occur if aerosols from one sample fall into another. Our equipment processes 8 samples simultaneously, and it is normal for customers to have doubts.


Why should we prevent such cross-contamination?

--Not all cross-contaminations have serious consequences, but cross-contamination in the molecular biology does produce extremely serious consequences because the back-end testing principle is multiple replication of the initial sample. Based on the current CT value of 35 as the cut-off value for a positive or negative diagnosis during the clinical test, the initial sample should be replicated 235 times, that is, 64.0 billion times. It is amazing that one nucleic acid can be amplified to 64.0 billion nucleic acids. Therefore, whenever a sample is contaminated by another one, result errors are inevitable.


2.Objective

To systematically evaluate the performance of YOCON automatic sample processing and nucleic acid extraction workstation and to verify that there is no aerosol contamination during normal extraction by the equipment.


3.Design

1、 A large number of strongly positive samples and negative samples were arranged in a staggered manner for replicate tests of 3 batches to test whether the extraction equipment would lead to cross-contamination under extreme conditions.

2、 A total of 48 samples were extracted for one batch, including 27 virus-containing samples and 21 virus-free samples.

3、 The virus concentration of the positive samples was 109 copies/mL (refer to the upper limit of actual concentration of positive clinical samples).

4、 Virus-containing samples and blank samples were arranged in a staggered manner, with positive samples around negative sample (negative-6 sample as shown below) and negative samples around positive sample (2-1 sample as shown below), as detailed below:


Arrangement of negative and positive samples.

Negative

Negative

Negative

Positive

Negative

Positive

Negative

Positive

Negative

Positive

Negative

Positive

Negative

Positive

Negative

Positive

Negative

Negative

Negative

Positive

Positive

Negative

Positive

Negative

Positive

Positive

Positive

Positive

Positive

Positive

Positive

Negative

Positive

Negative

Positive

Positive

Positive

Negative

Positive

Positive

Positive

Positive

Positive

Positive

Negative

Negative

Negative

Negative


4.Procedures

1、 The virus mixture was prepared.

2、 The virus samples were added. The virus mixture was added to the sampling tube in a 1:9 volume ratio. A total of 27 virus-containing sampling tubes were set, marked with serial numbers, and recorded. One swab was broken and put into each sampling tube. The tubes were shaken well and mixed well.

3、 Preparation of the reagents required for the instrument. A total of three reagents were required, that is, 15.36 mL of a mixture of isopropanol and magnetic beads (containing 14.4 mL of isopropanol and 960 μL of magnetic beads), 38.4 mL of purification solution and 4.8 mL of cleaning solution.

4、 27 tubes were inserted into the pre-set wells from left to right and top to bottom, and the negative control tube without virus sample were inserted into the remaining 21 wells to start the program for extraction.

5、 The reverse transcription fluorescence quantitative PCR assay was carried out for 21 negative samples, and one well containing viral nucleic acid was set as positive control.

6、 The replicate tests were carried out for 3 batches.


5.Results

1. The expansion curves of each group for batch 1 are as follows:

擴(kuò)增曲線1.jpg

Note: Only the positive control group had an expansion curve, while the negative control and others showed no expansion.


Statistics:

Group

Ct value

Group

Ct value

Group

Ct value

Negative-1

NoCt

Negative-8

NoCt

Negative-15

NoCt

Negative-2

NoCt

Negative-9

NoCt

Negative-16

NoCt

Negative-3

NoCt

Negative-10

NoCt

Negative-17

NoCt

Negative-4

NoCt

Negative-11

NoCt

Negative-18

NoCt

Negative-5

NoCt

Negative-12

NoCt

Negative-19

NoCt

Negative-6

NoCt

Negative-13

NoCt

Negative-20

NoCt

Negative-7

NoCt

Negative-14

NoCt

Negative-21

NoCt

Positive control

27.95

Negative control

NoCt




2. The expansion curves of each group for batch 2 are as follows:

擴(kuò)增曲線2.jpg

Note: Only the positive control group had an expansion curve, while the negative control and others showed no expansion.


Statistics:

Group

Ct value

Group

Ct value

Group

Ct value

Negative-1

NoCt

Negative-8

NoCt

Negative-15

NoCt

Negative-2

NoCt

Negative-9

NoCt

Negative-16

NoCt

Negative-3

NoCt

Negative-10

NoCt

Negative-17

NoCt

Negative-4

NoCt

Negative-11

NoCt

Negative-18

NoCt

Negative-5

NoCt

Negative-12

NoCt

Negative-19

NoCt

Negative-6

NoCt

Negative-13

NoCt

Negative-20

NoCt

Negative-7

NoCt

Negative-14

NoCt

Negative-21

NoCt

Positive control

28.55

Negative control

NoCt




3. The expansion curves of each group for batch 3 are as follows:

擴(kuò)增曲線3‘’.jpg

Note: Only the positive control group had an expansion curve, while the negative control and others showed no expansion.


Statistics:

Group

Ct value

Group

Ct value

Group

Ct value

Negative-1

NoCt

Negative-8

NoCt

Negative-15

NoCt

Negative-2

NoCt

Negative-9

NoCt

Negative-16

NoCt

Negative-3

NoCt

Negative-10

NoCt

Negative-17

NoCt

Negative-4

NoCt

Negative-11

NoCt

Negative-18

NoCt

Negative-5

NoCt

Negative-12

NoCt

Negative-19

NoCt

Negative-6

NoCt

Negative-13

NoCt

Negative-20

NoCt

Negative-7

NoCt

Negative-14

NoCt

Negative-21

NoCt

Positive control

29.72

Negative control

NoCt



6.Conclusions

During the extraction tests for the same batch, negative samples and positive samples were arranged in a staggered manner, and contamination was not found in negative sample wells, but found in positive control well, indicating that this equipment can effectively prevent inter-sample cross-contamination.


Words in the end:

We choose to be one of the few automatic nucleic acid workstation suppliers to offer measured data of samples using our automatic nucleic acid workstation. We will continue to test multiple batches of samples in the future to ensure that the equipment meets the customer needs. Let’s look forward to it!


一牛影视制作官方网站 | 免费无码又爽又高潮视频软件 | 国产成人三级一区二区在线观看一 | 丰满人妻精品乱子伦 | 三级片激情免费观看网站 | 91精品国产秘 入口在线 | 2024国产精品探花久久 | 91午夜理伦私人影院 | 安徽妇搡BBBB搡BBBB按摩 | 姝姝窝人体色www国产 | 人妻熟女中出在线4k | 国产伦子伦视频在线观看 | 91精品久久久久久久99蜜桃 | 91精品国产乱码久久久久 | 久久精品人妻一区二区三区宅男必备 | 人妻AⅤ无码一区二区三区 少妇无码免费在线A片免费 | 国产精品伦子伦免费 | 国产无码自拍视频 | 哔哩哔哩高清视频高清观看 | 少妇无套内谢太紧了A片软件 | 100大看免费行情的软件下载 | 秘书丝袜人妻中文字幕 | 欧美口爆视频在线播放 | 国产AV 无码 高潮 红桃 | 日本无码a午夜精品一区 | 国产一级婬乱片A片AAA图片 | 成人网站在线看污污污污 | 免费全黄A片免费播放 | 中文字幕在线中文幕免费在线看免费版 | 国产农村妇女一二三毛片 | 国产无遮挡A片又黄又爽小直播 | 日本成人电影中文字幕 | 四影虎影成人A片免费播放 日本有码性爱视频在线一区 | 国产成人精品无码免费播放精品 | 熟女人妻的婬乱和放荡 | 四川少妇渴BBBBB搡BBB | 成人做爰黄AA片免费看 | 免费白丝内内在线看 | 啪啪免费无码人妻丰满熟妇 | 新妺妺窝窝777777野外 | 一区二区三区精密机械公司 |